Photosynthesis apparatus meter资料参数

时间:2024-6-4阅读:410

Photosynthesis apparatus meter资料参数

Photosynthesis apparatus meter

一、Introduction to photosynthesis

1、A brief introduction to photosynthesis

1. Definition of photosynthetic rate (Pn)

The photosynthetic rate is the amount of CO2 fixed or O2 released or accumulated dry matter per unit time and photosynthetic mechanism (dry weight, area or chlorophyll). If CO2 is used as a measurement index, the unit of the photosynthetic rate is: μmolCO2m-2•s-1. This instrument uses the CO2 absorption of plant leaves to measure the photosynthetic rate of plant leaves.

2、Measurement principle of photosynthesis

Photosynthesis and transpiration are the internal and external exchange processes of CO2 and water vapor molecules through leaves respectively, and the main channels are stomata. The driving force of gas diffusion is the concentration difference between inside and outside the leaf. According to Fick's diffusion law, the diffusion flux of the gas in the diffusion path is directly proportional to the concentration difference of the diffusion path, and inversely proportional to the diffusion resistance. For the photosynthetic rate Pn, the relationship is as follows:


C —CO2 concentration difference at both ends of diffusion pathway;

It is the difference between the CO2 concentration in the air and the CO2 concentration in the mesophyll space

∑r—The total diffusion resistance of the diffusion pathway to CO2 molecules。

3、Measurement method

CO2 absorption method is divided into two measurement methods: closed circuit measurement and open circuit measurement.

二、 (Water Use Efficiency)

Water use efficiency (instantaneous water use efficiency) (wue) is equal

to the ratio of photosynthetic rate to transpiration rate:

三、 Transpiration

1、Definition of transpiration:

The loss of water in gaseous form through the surface of the plant to the outside is called transpiration, which is different from ordinary evaporation in that the former is the diffusion of water through the living organism, while the latter is a simple physical process. Transpiration is a physiological process, which is not only affected by environmental factors but also regulated by plant structure and stomatal behavior.

Transpiration rate definition:

The amount of water lost by transpiration per unit leaf area is called transpiration rate.

四、PAR( photosynthetically active radiation )

The radiation that plants use for photosynthesis in a specific wavelength range (400-700nm) is called photosynthetically active radiation.for short: PAR

五、Parameter annotation

Rni:  Humidity of the air before entering the leaf chamber

Rout: The humidity of the air after leaving the leaf chamber

RHc: The difference between Rout and Rni. RHc is used to calculate the transpiration rate

Flow: air flow rate

PARin: Photosynthetically active radiation sensor in the leaf chamber (with red and blue light source optional)

PARout: External photosynthetically active radiation sensor

CO2: The CO2 difference before and after the air passes through the leaf chamber

TC: leaf chamber temperature

TL: leaf temperature

RH: leaf chamber humidity

P: Atmospheric pressure

Pn: Photosynthetic rate

Tr: Transpiration rate

Wue: Water Use Efficiency

Gs: stomatal conductance

Ci: Intercellular CO2

六、Technical Parameters:

CO2 analysis:

A temperature-tuned dual-wavelength infrared carbon dioxide analyzer was added, measuring range: 0-3000 ppm, resolution: 0.1 ppm; accuracy 3 ppm. Carbon dioxide measurement is not affected by temperature changes, it is stable, high precision, and sensitive. The carbon dioxide difference can be collected within 1 second.

Leaf chamber temperature:

Germany Heraeus high-precision digital temperature sensor, measuring range: -20-80 ° C, resolution: 0.1 ° C, error ± 0.2 ° C

Blade temperature:

Platinum resistance, measuring range: -20-60 ° C, resolution: 0.1 ° C, error ± 0.2 ° C

humidity:

Swiss imported high precision digital humidity sensor:

Measuring range 0-100%, resolution: 0.1%, error ≤ 1%

Photosynthetically active radiation (PAR):

Silicon photocell with correction filter

Measuring range: 0-3000μmol m 2 / sec, accuracy<1μmolm m2 / sec. Response wavelength range: 400 ~ 700nm

Flow measurement: glass rotor flowmeter, flow rate is arbitrarily set within the range of 0-1.5L, error: 1%, <±0.2% in the range of 0.2~1L/min, the pump flow can be set according to needs, can measure different gases The effect of photosynthesis under flow, the gas flow is stable.

Leaf chamber size: standard size 55×20mm, other sizes are customized according to customer needs

Operating environment: temperature -20 ° C -60 ° C, relative humidity: 0-100% (no condensation)

Power supply: DC8.4V lithium battery, can work continuously for 10 hours

Data storage: 16G memory, expandable to 32G.

Data transfer: USB connected to the computer can directly export data.

Display: 3.5" TFT true color LCD color display, resolution 800 × 480, clearly visible under strong light

Volume: 260 × 260 × 130mm


会员登录

X

请输入账号

请输入密码

=

请输验证码

收藏该商铺

X
该信息已收藏!
标签:
保存成功

(空格分隔,最多3个,单个标签最多10个字符)

常用:

提示

X
您的留言已提交成功!我们将在第一时间回复您~

以上信息由企业自行提供,信息内容的真实性、准确性和合法性由相关企业负责,化工仪器网对此不承担任何保证责任。

温馨提示:为规避购买风险,建议您在购买产品前务必确认供应商资质及产品质量。

拨打电话
在线留言
渝中区| 梓潼县| 福贡县| 南阳市| 天全县| 安阳县| 吉木萨尔县| 台湾省| 涞水县| 兴海县| 自贡市| 新龙县| 柘城县| 双峰县| 探索| 丰原市| 卫辉市| 织金县| 南安市| 大渡口区| 克什克腾旗| 四子王旗| 锦州市| 巴南区| 义马市| 繁峙县| 盘山县| 丹棱县| 内乡县| 新昌县| 四川省| 巴彦县| 波密县| 始兴县| 钟山县| 孟州市| 延庆县| 芦山县| 昆山市| 新沂市| 梅河口市|